首页 | 本学科首页   官方微博 | 高级检索  
     

国产医用直线加速器VenusX稳定性分析
引用本文:解传滨,申红峰,戴相昆,高文超,王军良,刘彦立,俞伟,曲宝林. 国产医用直线加速器VenusX稳定性分析[J]. 中国医疗设备, 2021, 0(4): 61-64,75
作者姓名:解传滨  申红峰  戴相昆  高文超  王军良  刘彦立  俞伟  曲宝林
作者单位:中国人民解放军总医院第一医学中心放射治疗科;中国人民解放军总医院第五医学中心放射治疗科
基金项目:国家重点研发计划(2017YFC0112100);军队保健课题(20BJZ40)。
摘    要:目的 通过对国产医用直线加速器VenusX执行为期3个月的质控监测,分析其机械精度、射束性能和影像系统的稳定性,为其下一步临床应用推广提供数据参考.方法 基于VenusX的EPID系统,应用DVS系统对加速器的机械精度、射束性能和影像系统等进行质量检测,并给出相应质检报告,分析评价该型加速器的稳定性和可靠性.结果 Ve...

关 键 词:医用直线加速器  质量控制  准确性  稳定性

Stability Analysis for Domestic Medical Linear Accelerator VenusX
XIE Chuanbin,SHEN Hongfeng,DAI Xiangkun,GAO Wenchao,WANG Junliang,LIU Yanli,YU wei,QU Baolin. Stability Analysis for Domestic Medical Linear Accelerator VenusX[J]. Chinese medical equipment, 2021, 0(4): 61-64,75
Authors:XIE Chuanbin  SHEN Hongfeng  DAI Xiangkun  GAO Wenchao  WANG Junliang  LIU Yanli  YU wei  QU Baolin
Affiliation:(Department of Radiotherapy,The First Medical Center,Chinese PLA General Hospital,Beijing 100853,China;Department of Radiotherapy,The Fifth Medical Center,Chinese PLA General Hospital,Beijing 100853,China)
Abstract:Objective To analyse the stability of mechanical accuracy,beam performance and image system for domestic medical linear accelerator VenusX based on a three-month quality control monitoring,and provide data reference for the next clinical application promotion.Methods The quality inspections of mechanical accuracy,beam performance and image system for the VenusX were carried out and the corresponding reports were gived based on the EPID and DVS system.The Stability and Reliability of the accelerator were analyzed and evaluated.Results The stability of the accelerator parameters were in accordance with the corresponding standards.The maximum deviations of X-ray output constancyand deviation of doserate stability were 0.5%and 1%.The maximum deviations of gantry rotation was 0.8 mm.The maximum position accuracy of MLC was 1 mm,The maximum repeatability deviations of MLC in the X and Y direction were 0.3 mm and 0.4 mm.The maximum deviations of central consistency of MV-EPID and KV-EPID were 0.5 mm and 0.6 mm,respectively.Conclusion The mechanical accuracy,beam performance and image system stability of VenusX were in line with the relevant national standards and met the clinical requirements according to the daily stability analysis.
Keywords:medical linear accelerator  quality control  accuracy  reliability
本文献已被 维普 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号