Spiral readout gradients for the reduction of motion artifacts in chemical shift imaging. |
| |
Authors: | Dong-Hyun Kim Elfar Adalsteinsson Daniel M Spielman |
| |
Affiliation: | Radiological Sciences Laboratory, Department of Radiology, Stanford University, Stanford, California, USA. dhkim@stanford.edu |
| |
Abstract: | A motion artifact reduction method for proton chemical shift imaging (CSI) is presented. The method uses spiral-based readout gradients for data acquisition. A characteristic of spiral-based readout gradients is that data are repeatedly sampled at the kxy origin. These data points are used to estimate and correct for motion-induced phase variations. Both phantom and in vivo spectra reconstructed using the new motion artifact reduction algorithm showed significant signal-to-noise ratio (SNR) improvements as compared to uncorrected data. |
| |