首页 | 本学科首页   官方微博 | 高级检索  
     


The Harris-Galante total hip arthroplasty: a minimum 8-year follow-up study
Authors:Cruz-Pardos A  Garcia-Cimbrelo E
Affiliation:Orthopaedic Department, Hospital la Paz, Plaza Reyes Magos 4, 28007 Madrid, Spain.
Abstract:A total of 93 Harris-Galante Porous I (HGP- I) prostheses were analyzed with a mean follow-up of 10.2 years. Thigh pain (28 hips) correlated with unstable fixation (P<.005), female gender (P<.045), young age (P<.05), poor distal femoral fill (P<.002), subsidence (P<.0001), and osteolysis (P<.05). Thirteen stems and 6 metallic shell cups were revised. Kaplan-Meier analysis showed 13-year survival rates of 92.5% +/- 5.8% for the metallic shell cup, 79.7% +/- 13% for the HGP-I cup, and 76.3% +/- 14.0% for the stem. Five polyethylene liners were changed because of wear >1 mm. Radiographic loosening occurred in 8 cups. Radiographic bone ingrowth was present in 54 stems, stable fibrous fixation was present in 24 stems, and unstable fixation was present in 15 stems. The mean polyethylene wear was 0.17 mm/y. Eleven hips (11.8%) had acetabular osteolysis, and 24 (25.8%) had femoral osteolysis, the latter being more frequent in unstable stems (P<.007). The HGP-I metallic cup shows better clinical and radiographic results than the stem, which frequently is associated with pain, unstable fixation, and osteolysis. Excessive polyethylene wear is frequent.
Keywords:
本文献已被 ScienceDirect PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号