Rapid exclusion of the diagnosis of immune HIT by AcuStar HIT and heparin-induced multiple electrode aggregometry
Authors:V Minet  J Baudar  N Bailly  J Douxfils  J Laloy  S Lessire  M Gourdin  B Devalet  B Chatelain  JM Dogné  F Mullier
Institution:1. Department of Pharmacy, Namur Thrombosis Hemostasis Center (NTHC), Namur Research Institute for LIfe Sciences (NARILIS), University of Namur, Namur, Belgium;2. Hematology Laboratory, Namur Thrombosis and Hemostasis Center (NTHC), Namur Research Institute for LIfe Sciences (NARILIS), CHU Dinant Godinne - UCL Namur, Belgium;3. Anesthesiology Department, Namur Thrombosis and Hemostasis Center (NTHC), CHU Dinant Godinne - UCL Namur, Belgium;4. Hematology Department, Namur Thrombosis Hemostasis Center (NTHC), Namur Research Institute for LIfe Sciences (NARILIS), CHU Dinant Godinne - UCL Namur, Belgium
Abstract:

Background

Accurate diagnosis of heparin-induced thrombocytopenia (HIT) is essential but remains challenging. We have previously demonstrated, in a retrospective study, the usefulness of the combination of the 4Ts score, AcuStar HIT and heparin-induced multiple electrode aggregometry (HIMEA) with optimized thresholds.

Objectives

We aimed at exploring prospectively the performances of our optimized diagnostic algorithm on suspected HIT patients. The secondary objective is to evaluate performances of AcuStar HIT-Ab (PF4-H) in comparison with the clinical outcome.

Methods

116 inpatients with clinically suspected immune HIT were included. Our optimized diagnostic algorithm was applied to each patient. Sensitivity, specificity, negative predictive value (NPV), positive predictive value (PPV) of the overall diagnostic strategy as well as AcuStar HIT-Ab (at manufacturer’s thresholds and at our thresholds) were calculated using clinical diagnosis as the reference.

Results

Among 116 patients, 2 patients had clinically-diagnosed HIT. These 2 patients were positive on AcuStar HIT-Ab, AcuStar HIT-IgG and HIMEA. Using our optimized algorithm, all patients were correctly diagnosed. AcuStar HIT-Ab at our cut-off (> 9.41 U/mL) and at manufacturer’s cut-off (> 1.00 U/mL) showed both a sensitivity of 100.0% and a specificity of 99.1% and 90.4%, respectively.

Conclusion

The combination of the 4Ts score, the HemosIL® AcuStar HIT and HIMEA with optimized thresholds may be useful for the rapid and accurate exclusion of the diagnosis of immune HIT.
Keywords:immune heparin-induced thrombocytopenia  HemosIL AcuStar HIT  heparin-induced multiple electrode aggregometry