床旁监测血糖仪葡萄糖氧化酶法及脱氢酶法准确度评估
引用本文:邸玉玮,郑卫东,梁雪冰,戴耀宗.床旁监测血糖仪葡萄糖氧化酶法及脱氢酶法准确度评估[J].中华内分泌代谢杂志,2011,27(8).
作者姓名:邸玉玮  郑卫东  梁雪冰  戴耀宗
作者单位:广东省医学科学院,广东省人民医院检验科,广州,510018
摘    要:以生化分析仪检测为参照,比较葡萄糖氧化酶法及脱氢酶法血糖仪的准确度.检测新鲜肝素锂化静脉血临床样本.根据ISO15197,血糖仪的两种方法检测结果偏倚100%在±0.56 mmol/L内(<4.2mmol/L),≥99.7%在±20%内(≥4.2 mmol/L);根据EF9-A2,两种方法的预期误差<10%,略>5%;低、高血糖症界值处(<3.89 mmol/L或>6.11mmol/L),两种方法的κ>0.6.两种方法间的κ<0.6.血糖仪葡萄糖氧化酶法、脱氢酶法准确度符合ISO15197要求,与生化分析仪检测方法的差异不影响临床应用,其结果用于低、高血糖症判读一致性好.
Abstract:
Blood samples were detected by 24 blood-glucose meters using glucose oxidase and another 18 glucose meters using glucose dehydrogenase. The plasma glucose was detected by the auto-chemistry analyzer as control. According to ISO1 5197 and EF9-A2, the bias of results from both glucose meters 100% fall in the range of ±0. 56 mmol/L( <4. 2 mmol/L) and ≥99. 7% in the range of ±20% ( ≥4. 2 mmol/L), predicted bias were all less than the true bias(Bc). As κ>0. 6, the results from both glucose meters were in accordance with the results from autochemistry analyzer in judging hyperglycemia ( >6. 11 mmol/L) and hypoglycemia ( < 3.89 mmol/L). Between the results from the two blood-glucose meters, κ<0. 6. The accuracy of both glucose-meters are accepted for the purpose of clinical diagnosis and treatment.

关 键 词:血糖  测定  准确性

Accuracy of blood-glucose monitoring system by using glucose oxidase or glucose dehydrogenase
DI Yu-wei,ZHENG Wei-dong,LIANG Xue-bing,DAI Yao-zhong.Accuracy of blood-glucose monitoring system by using glucose oxidase or glucose dehydrogenase[J].Chinese Journal of Endocrinology and Metabolism,2011,27(8).
Authors:DI Yu-wei  ZHENG Wei-dong  LIANG Xue-bing  DAI Yao-zhong
Abstract:Blood samples were detected by 24 blood-glucose meters using glucose oxidase and another 18 glucose meters using glucose dehydrogenase. The plasma glucose was detected by the auto-chemistry analyzer as control. According to ISO1 5197 and EF9-A2, the bias of results from both glucose meters 100% fall in the range of ±0. 56 mmol/L( <4. 2 mmol/L) and ≥99. 7% in the range of ±20% ( ≥4. 2 mmol/L), predicted bias were all less than the true bias(Bc). As κ>0. 6, the results from both glucose meters were in accordance with the results from autochemistry analyzer in judging hyperglycemia ( >6. 11 mmol/L) and hypoglycemia ( < 3.89 mmol/L). Between the results from the two blood-glucose meters, κ<0. 6. The accuracy of both glucose-meters are accepted for the purpose of clinical diagnosis and treatment.
Keywords:Blood glucose  Measurement  Accuracy