An operative case of inferior vena cava stenosis due to fibrosis around permanent pacemaker leads
Authors:Shimada Yasuyuki  Yaku Hitoshi  Kawata Masatoshi  Oka Katsuhiko  Shuntoh Keisuke  Okano Takahisa  Takahashi Akiyuki  Fukumoto Atsushi  Hayashida Kyoko  Kitamura Nobuo
Institution:Department of Cardiovascular Surgery, Kyoto Prefectural University of Medicine, Kawaramachi, Japan. Yasuyuki.Shimada@ma8.seikyou.ne.jp
Abstract:A 56-year-old woman presented with general fatigue, dyspnea on exercise, and hepatomegaly subsequent to secondary implantation of a pacemaker. On admission, cardiac echo examination showed tricuspid valve regurgitation due to a migrated looped lead. At surgery, there was evidence of severe stenosis at both orifices of the superior and inferior vena cavae due to fibrous tissue around the leads. We removed the fibrous tissue, pacing leads, and generator. New leads were anchored onto the epicardium and a generator was inserted under the rectus muscle. The postoperative course was satisfactory and the symptoms disappeared.
Keywords:vena cava  stenosis  pacemaker  complication