首页 | 本学科首页   官方微博 | 高级检索  
     


Measurement of tidal volume using a pneumotachometer during high-frequency oscillation
Authors:S E Courtney  K R Weber  W A Spohn  S W Malin  C V Bender  R W Gotshall
Affiliation:Department of Pediatrics, Wright State University, Dayton, OH 45435.
Abstract:To assess the accuracy of a pneumotachometer (PN) for tidal volume (VT) measurements during high-frequency oscillation (HFO), we determined simultaneously VT using a PN and a full body plethysmograph (PL) in 12 rabbits. HFO was delivered with an oscillator at a frequency of 10 Hz, mean airway pressure of 8 cm H2O, and inspiratory time of 50%. Pressure amplitude (delta P) was varied as follows: 40, 60, 80, 20, 100, 40 cm H2O. Finally, in ten rabbits a spacer equal in deadspace (VD) to that of the PN (15 ml) was left in-line for 5 min. Blood gases were obtained before and after the spacer was added. We found that VT-PN correlates well with VT-PL (r = .92), although the difference between VT-PN and VT-PL is greater at large VT. Significant respiratory acidosis developed with the spacer in-line. PN may be used to trend VT during HFO but PN must not be left in-line, as increased VD seriously affects ventilation.
Keywords:
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号