Parallel MRI with extended and averaged GRAPPA kernels (PEAK-GRAPPA): optimized spatiotemporal dynamic imaging
Authors:Jung Bernd  Ullmann Peter  Honal Matthias  Bauer Simon  Hennig Jürgen  Markl Michael
Institution:Department of Diagnostic Radiology, Medical Physics, University Hospital, Freiburg, Germany. bernd.jung@uniklinik-freiburg.de
Abstract:

Purpose

To evaluate an optimized k‐t‐space related reconstruction method for dynamic magnetic resonance imaging (MRI), a method called PEAK‐GRAPPA (Parallel MRI with Extended and Averaged GRAPPA Kernels) is presented which is based on an extended spatiotemporal GRAPPA kernel in combination with temporal averaging of coil weights.

Materials and Methods

The PEAK‐GRAPPA kernel consists of a uniform geometry with several spatial and temporal source points from acquired k‐space lines and several target points from missing k‐space lines. In order to improve the quality of coil weight estimation sets of coil weights are averaged over the temporal dimension.

Results

The kernel geometry leads to strongly decreased reconstruction times compared to the recently introduced k‐t‐GRAPPA using different kernel geometries with only one target point per kernel to fit. Improved results were obtained in terms of the root mean square error and the signal‐to‐noise ratio as demonstrated by in vivo cardiac imaging.

Conclusion

Using a uniform kernel geometry for weight estimation with the properties of uncorrelated noise of different acquired timeframes, optimized results were achieved in terms of error level, signal‐to‐noise ratio, and reconstruction time. J. Magn. Reson. Imaging 2008;28:1226–1232. © 2008 Wiley‐Liss, Inc.
Keywords:parallel MRI  GRAPPA  dynamic imaging