Measurement of liver stiffness with two imaging techniques: magnetic resonance elastography and ultrasound elastometry |
| |
Authors: | Bensamoun Sabine F Wang Lu Robert Ludovic Charleux Fabrice Latrive Jean-Paul Ho Ba Tho Marie-Christine |
| |
Affiliation: | Biomechanics and Bioengineery Laboratory, UMR CNRS 6600, Université de Technologie de Compiègne, and Centre hospitalier de Compiègne, service gastro-entérologie et hépatologie, Compiègne, France. sabine.bensamoun@utc.fr |
| |
Abstract: | Purpose To cross‐validate the magnetic resonance elastography (MRE) technique with a clinical device, based on an ultrasound elastometry system called Fibroscan. Materials and Methods Ten healthy subjects underwent an MRE and a Fibroscan test. The MRE technique used a round pneumatic driver at 60 Hz to generate shear waves inside the liver. An elastogram representing a map of the liver stiffness was generated allowing for the measurement of the average liver stiffness inside a region of interest. The Fibroscan technique used an ultrasound probe (3.5 MHz) composed of a vibrator that sent low‐frequency (50 Hz) shear waves inside the right liver lobe. The probe acts as an emitter‐receptor that measures the velocity of the waves propagated inside the liver tissue. Results The mean shear stiffness measured with the MRE and Fibroscan techniques were 1.95 ± 0.06 kPa and 1.79 ± 0.30 kPa, respectively. A higher standard deviation was found for the same subject with Fibroscan. Conclusion This study shows why MRE should be investigated beyond the Fibroscan. The MRE technique provided elasticity of the entire liver, meanwhile the Fibroscan provided values of elasticity locally. J. Magn. Reson. Imaging 2008;28:1287–1292. © 2008 Wiley‐Liss, Inc. |
| |
Keywords: | magnetic resonance elastography Fibroscan liver cross‐validation |
本文献已被 PubMed 等数据库收录! |
|