阿洛卡SSD-2000超声仪故障检修分析
引用本文:赵武. 阿洛卡SSD-2000超声仪故障检修分析[J]. 中国医学装备, 2008, 5(5): 64-66
作者姓名:赵武
作者单位:内蒙古赤峰市医院设备科,内蒙古,赤峰,024000
摘    要:目的:对阿洛卡超声仪无图像或图像时有时无的故障检修。方法:采用对怀疑电容加入直流5V电压的方法,可快速准确的找到故障电容。结论:在维修中应首先从电源入手,测量其各路电压是否正常。在对有故障机器的检修后总结发现,电源的故障率在整台机器的故障中可达到40%以上,这一点在维修中应引起工程师的高度重视,有时可起到事半功倍的效果。

关 键 词:电源  电压  电容  变压器
文章编号:1672-8270(2008)05-0064-03
修稿时间:2008-02-18

To solve imaging problems of Aloka Ultrasound Instrument
Zhao Wu. To solve imaging problems of Aloka Ultrasound Instrument[J]. China Medical Equipment, 2008, 5(5): 64-66
Authors:Zhao Wu
Affiliation:Zhao Wu (Chifeng Hospital of Neimenggu,Neimenggu 024000,China)
Abstract:Purpose: to solve imaging problems of Aloka Ultrasound Instrument. Methods: we can find the capacitance with fault quickly and accurately through joining 5 volts DC to suspected capacitance. Conclusion: During repairing, check the power first, and then measure the voltages of each pathway. We made a conclusion that the error rate of the power in the whole machine reaches up to 40%. Engineers should pay great attention to power error, sometimes this can make our work easily and fast.