Estimation of personal dose based on the dependent calibration of personal dosimeters in interventional radiology
Authors:Mori Hiroshige  Koshida Kichiro  Ichikawa Katsuhiro
Affiliation:Department of Radiology, Hokkaido Social Insurance Hospital.
Abstract:The purpose of present study is, in interventional radiology (IVR), to elucidate the differences between each personal dosimeter, and the dependences and calibrations of area or personal dose by measurement with electronic dosimeters in particular. We compare space dose rate distributions measured by an ionization survey meter with the value measured by personal dosimeter: an optically stimulated luminescence, two fluoroglass, and two electronic dosimeters. Furthermore, with electronic dosimeters, we first measured dose rate, energy, and directional dependences. Secondly, we calibrated the dose rate measured by electronic dosimeters with the results, and estimated these methods with coefficient of determination and Akaike's Information Criterion (AIC). The results, especially in electronic dosimeters, revealed that the dose rate measured fell by energy and directional dependences. In terms of methods of calibration, the method is sufficient for energy dependence, but not for directional dependence, because of the lack of stable calibration. This improvement poses a question for the future. The study suggested that these dependences of the personal dosimeter must be considered when area or personal dose is estimated in IVR.
Keywords: