Comparison of the Cobas-Bact five-hour susceptibility testing system with the NCCLS agar diffusion and dilution methods |
| |
Authors: | J Würst W Heizmann U Hardegger B Manncke |
| |
Institution: | (1) Department of Medical Microbiology, University of Zürich, Gloriastraße 32, 8028 Zürich, Switzerland;(2) Department of Medical Microbiology, University of Tübingen, Silcherstraße 7, 7400 Tübingen, FRG |
| |
Abstract: | The results of susceptibility tests performed by the Cobas-Bact system were compared with those of the NCCLS agar diffusion (Kirby-Bauer) and NCCLS agar dilution methods. A total of 998 clinical isolates were tested against 10 to 18 antimicrobial agents. Essential agreement (comprising full agreement and minor discrepancies) varied from 90.5 % to 99.2 % on comparison of Cobas-Bact with Kirby-Bauer results, depending on the bacterial group (mean for all 998 strains tested 95.7 %). These figures ranged from 91 % to 99.2 % (mean 96.3 %) for the Cobas-Bact/MIC comparison and from 95.2 % to 99.7 % (mean 98.7 %) for the Kirby-Bauer/MIC comparison. The best results were found forEnterobacteriaceae andStaphylococcus aureus, whereas for enterococci and coagulase-negative staphylococci there was a lower rate of essential agreement in all three comparisons. In the case ofPseudomonas aeruginosa there was a good rate of essential agreement but many minor discrepancies, resulting in a disappointing rate of full agreement of between 67.5 % and 78.9 % in the three comparisons. The Cobas-Bact system would appear to provide satisfactory susceptibility test results in most cases, however there are still some major problems in the system which should be resolved. |
| |