首页 | 本学科首页   官方微博 | 高级检索  
检索        


The relationship between stress and relaxation in calf pericardium used in the construction of cardiac bioprostheses
Authors:J M García Páez  A Carrera San Martín  J V García Sestafe  E Jorge  I Millán  I Candela
Institution:Experimental Surgery Service, Clinica Puerta de Hierro, Spain.
Abstract:Assessment of relaxation (loss of load within a given time) without apparent deformation is a necessary step before durability assay of biomaterials. From results obtained using calf pericardium, the following conclusions were drawn: (a) there is no limit to relaxation for this biomaterials; (b) the lesser the load applied, the greater the relaxation; and (c) the relaxation curve follows the logarithmic function y = K1 - K2 - Int. These findings demonstrate the poor behaviour of the biomaterial at supposedly low loads and suggest that rupture-point load is not a reliable reference to determine the safety coefficient of calf pericardium used in cardiac bioprostheses.
Keywords:
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号