首页 | 本学科首页   官方微博 | 高级检索  
     


Persistently better treatment planning results of intensity-modulated (IMRT) over conformal radiotherapy (3D-CRT) in prostate cancer patients with significant variation of clinical target volume and/or organs-at-risk
Authors:Pascal Fenoglietto   Benoit Laliberte   Ali Allaw   Norbert Ailleres   Katia Idri   Meng Huor Hay   Carmen Llacer Moscardo   Sophie Gourgou   Jean-Bernard Dubois  David Azria
Affiliation:Département d'Oncologie Radiothérapie, CRLC Val d'Aurelle-Paul Lamarque, Montpellier, France. pfenoglietto@valdorel.fnclcc.fr
Abstract:PURPOSE: To compare the dose coverage of planning and clinical target volume (PTV, CTV), and organs-at-risk (OAR) between intensity-modulated (3D-IMRT) and conventional conformal radiotherapy (3D-CRT) before and after internal organ variation in prostate cancer. METHODS AND MATERIALS: We selected 10 patients with clinically significant interfraction volume changes. Patients were treated with 3D-IMRT to 80 Gy (minimum PTV dose of 76 Gy, excluding rectum). Fictitious, equivalent 3D-CRT plans (80 Gy at isocenter, with 95% isodose (76 Gy) coverage of PTV, with rectal blocking above 76 Gy) were generated using the same planning CT data set ("CT planning"). The plans were then also applied to a verification CT scan ("CT verify") obtained at a different moment. PTV, CTV, and OAR dose coverage were compared using non-parametric tests statistics for V95, V90 (% of the volume receiving 95 or 90% of the dose) and D50 (dose to 50% of the volume). RESULTS: Mean V95 of the PTV for "CT planning" was 94.3% (range, 88-99) vs 89.1% (range, 84-94.5) for 3D-IMRT and 3D-CRT (p=0.005), respectively. Mean V95 of the CTV for "CT verify" was 97% for both 3D-IMRT and 3D-CRT. Mean D50 of the rectum for "CT planning" was 26.8 Gy (range, 22-35) vs 43.5 Gy (range, 33.5-50.5) for 3D-IMRT and 3D-CRT (p=0.0002), respectively. For "CT verify", this D50 was 31.1 Gy (range, 16.5-44) vs 44.2 Gy (range, 34-55) for 3D-IMRT and 3D-CRT (p=0.006), respectively. V95 of the rectum was 0% for both plans for "CT planning", and 2.3% (3D-IMRT) vs 2.1% (3D-CRT) for "CT verify" (p=non-sig.). CONCLUSION: Dose coverage of the PTV and OAR was better with 3D-IMRT for each patient and remained so after internal volume changes.
Keywords:Prostate cancer   Intensity-modulated radiotherapy   Conformal radiotherapy   Organ motion   Treatment planning
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号