Application of Computed Tomography-Based Navigation for Revision Total Hip Arthroplasty |
| |
Authors: | Nobuo Nakamura Takashi Nishii Makoto Kitada Daiki Iwana Nobuhiko Sugano |
| |
Institution: | 1. Center of Arthroplasty, Kyowakai Hospital, Suita-city, Osaka, Japan;2. Department of Orthopedic Surgery, Osaka University Medical School, Suita-city, Osaka, Japan |
| |
Abstract: | The usefulness of navigation systems for revision total hip arthroplasty (THA) remains unclear. The purpose of this study was to evaluate the accuracy of cup positioning using a navigation system in revision THA. Cup alignment of thirty revision cases was evaluated. Data were also compared with those of 30 navigated primary THA cases. In the revision THA group, mean difference in postoperative measurement from preoperative plan was 0.5 ± 4° (range, − 9° to 8°) for inclination and − 0.07° ± 5° (− 14° to 13°) for anteversion. Mean difference in postoperative measurement from intraoperative record was 0.7° ± 3° (− 5° to 8°) for inclination and − 0.7° ± 3° (− 12° to 5°) for anteversion. Compared with the primary THA group, none of these parameters were significantly different. This navigation system was useful even in revision THA. |
| |
Keywords: | revision total hip arthroplasty CT-based navigation metal implant |
本文献已被 ScienceDirect 等数据库收录! |
|