压力蒸汽灭菌用自动启闭式盛装容器性能测定
引用本文:姚楚水,刘万鹏,刘育京.压力蒸汽灭菌用自动启闭式盛装容器性能测定[J].中国消毒学杂志,1988(4).
作者姓名:姚楚水  刘万鹏  刘育京
作者单位:军事医学科学院微生物流行病研究所 (姚楚水),空军后勤部门诊部 (刘万鹏),军事医学科学院微生物流行病研究所(刘育京)
摘    要:<正> 医院中的注射器采用压力蒸汽灭菌法处理时,若所用盛装容器不当,往往造成灭菌失败。普通铝饭盒透气性差,不适合做为注射器在压力蒸汽灭菌处理时的盛装容器,但目前国内医院中还普遍使用。为了提供一种适宜并使用方便的注射器盛装容器进行


EXAMINATION OF PERFORMANCE OF AUTO-OPEN-CLOSE CONTAINER FOR AUTOCLAVING
Abstract:A new type of auto-open-close container for autoclaving of syringes was developed. On the lower part of lateral sides of the container, there are air holes which can open automatically to facilitate replacement of air in the container when the temperature rises above 50℃. After sterilization, the air holes can close automatically when the temperature reduces to 50℃ to prevent recontamination. By examination, it was found that all the syringes in the container could be sterilized satisfactorily after autoclaving at 121℃ for 30 min(with Bacillus stearothermophilus as the biological indicator). No recontamination of sterilized syringes in the container was found after storage for 2 weeks.