Hyphema caused by a metallic intraocular foreign body during magnetic resonance imaging
Authors:Ta C N  Bowman R W
Institution:Department of Ophthalmology, Stanford University Medical Center, Stanford, California 94305-5308, USA.
Abstract:PURPOSE: To report a 63-year-old man with a retained intraocular foreign body who developed a hyphema during magnetic resonance imaging (MRI) of the brain. METHODS: Case report and review of the current literature on ocular injury caused by intraocular foreign bodies when subjected to an electromagnetic field. RESULTS: Our patient underwent a brain MRI, and the intraocular foreign body caused a hyphema and increased intraocular pressure. The presence and location of the intraocular foreign body were determined by computed tomography (CT). CONCLUSION: Magnetic resonance imaging can cause serious ocular injury in patients with ferromagnetic intraocular foreign bodies. This case demonstrates the importance of obtaining an occupational history, and, when indicated, a skull x-ray or CT to rule out intraocular foreign body before an MRI study.