Hyphema caused by a metallic intraocular foreign body during magnetic resonance imaging |
| |
Authors: | Ta C N Bowman R W |
| |
Institution: | Department of Ophthalmology, Stanford University Medical Center, Stanford, California 94305-5308, USA. |
| |
Abstract: | PURPOSE: To report a 63-year-old man with a retained intraocular foreign body who developed a hyphema during magnetic resonance imaging (MRI) of the brain. METHODS: Case report and review of the current literature on ocular injury caused by intraocular foreign bodies when subjected to an electromagnetic field. RESULTS: Our patient underwent a brain MRI, and the intraocular foreign body caused a hyphema and increased intraocular pressure. The presence and location of the intraocular foreign body were determined by computed tomography (CT). CONCLUSION: Magnetic resonance imaging can cause serious ocular injury in patients with ferromagnetic intraocular foreign bodies. This case demonstrates the importance of obtaining an occupational history, and, when indicated, a skull x-ray or CT to rule out intraocular foreign body before an MRI study. |
| |
Keywords: | |
本文献已被 PubMed 等数据库收录! |
|