两种计算正常成年人短时心率变异性方法的比较研究
引用本文:李静,王步青,王卫东. 两种计算正常成年人短时心率变异性方法的比较研究[J]. 北京生物医学工程, 2017, 36(5). DOI: 10.3969/j.issn.1002-3208.2017.05.011
作者姓名:李静  王步青  王卫东
作者单位:中国人民解放军总医院医学工程与维修中心 北京100853;中国人民解放军总医院医学工程与维修中心 北京100853;中国人民解放军总医院医学工程与维修中心 北京100853
基金项目:国家自然科学基金,国家科技支撑计划
摘    要:目的利用光电容积脉搏波法分析心率变异性,为通过心率变异性(heart rate variability,HRV)参数表征自主神经系统的变化提供新的研究方法。方法对46个健康成年受试者分别同步采集自主呼吸状态下的心电信号和光电容积脉搏波信号,利用二阶差分极大值提取心电图的RR间期,搜索脉搏波极大值提取脉搏波主波波峰间期PP,然后计算通用HRV参数SDNN、RMSSD和LF/HF,并利用回归的统计学方法结合Bland-Altman随机分析法对这两种方法获取的HRV结果进行比较。结果两种方法计算得到的心率变异性参数结果的Pearson相关系数分别为0.998、0.995和0.992,均显著相关(P0.01),Bland-Altman分布图也均在一致性区间内,说明这两种方法在计算心率变异性上具有一致性。结论利用光电容积脉搏波法分析心率变异具有可行性。

关 键 词:心电图  光电容积脉搏波  心率变异性  一致性

Comparative study on the measurement of short-time heart rate variability using two methods in healthy adults
LI Jing,WANG Buqing,WANG Weidong. Comparative study on the measurement of short-time heart rate variability using two methods in healthy adults[J]. Beijing Biomedical Engineering, 2017, 36(5). DOI: 10.3969/j.issn.1002-3208.2017.05.011
Authors:LI Jing  WANG Buqing  WANG Weidong
Abstract:Objective To study the measurement of heart rate variability by using electrocardiogram ( ECG) and photoplethysmography( PPG) . Methods We analyzed heart rate variability ( HRV) taken by ECG and photoplethysmography. After acquiring the synchronized ECG as well as the photoplethysmography on 46 healthy adults in their natural breathing state and extracting the RR intervals of ECG and PP intervals of PPG respectively,the time domain and frequency domain parameters of HRV were calculated. Results The correlation coefficients of the three short-time parameters in the two calculation methods were greater than 0. 99 significantly,and the Bland-Altman scattered plot of the parameters were in the consistent band. The two methods had coherence in the calculation of heart rate variability. Conclusions It is feasible to analyze the heart rate variation with the pulse wave method.
Keywords:electrocardiogram  photoplethysmography  heart rate variability  coherence