首页 | 本学科首页   官方微博 | 高级检索  
检索        


Bridge to transplantation with the Jarvik-7 (CardioWest) total artificial heart: a single-center 15-year experience
Authors:P Leprince MD  N Bonnet MD  A Rama MD  P Lger MD  V Bors MD  J P Levasseur MD  J Szefner MD  E Vaissier MD  A Pavie MD  I Gandjbakhch MD
Institution:Department of Cardiothoracic Surgery, La Pitié-Salpétrière Hospital, Paris, France. pascal.leprince@psl.ap-hop-paris.fr
Abstract:BACKGROUND: At our institution, the total artificial heart (TAH) Jarvik-7 (CardioWest) has been used since 1986 as a bridge to transplantation for the most severely ill patients with terminal congestive heart failure. METHODS: Between 1986 and 2001, 127 patients (108 males, mean age 38 +/- 13) were bridged to transplantation with the Jarvik-7 TAH. All were in terminal biventricular failure despite high-dose inotropic support. Nine patients had a body surface area (BSA) of <1.6 m(2). In Group I patients (78%), the etiology of cardiac failure was dilated cardiomyopathy, either idiopathic (n = 60) or ischemic (n = 38). The other 29 patients (Group II) had disease of miscellaneous origin. We analyzed our experience with regard to 3 time periods: 1986 to 1992 (n = 63); 1993 to 1997 (n = 36); and 1998 to 2001 (n = 33). RESULTS: Although Group II patients represented 30% of indications before 1992, they comprised only 15% during the 2 subsequent periods. Duration of support for transplant patients increased dramatically after 1997, reaching 2 months for the most recent period (5 to 271 days). In Group I, the percentage of transplanted patients increased from 43% before 1993 to 55% between 1993 and 1997, and reached 74% thereafter. The major cause of death was multiorgan failure (67%). The clinical thromboembolic event rate was particularly low with no instance of cerebrovascular accident and 2 transient ischemic attacks. Total bleeding complication rate was 26%, including 2 deaths related to intractable hemorrhage and 2 others related to atrial tamponade. The cumulative experience was 3,606 total implant days with only 1 instance of mechanical dysfunction. CONCLUSIONS: TAH is a safe and efficient bridge for patients with terminal congestive heart failure awaiting cardiac transplantation.
Keywords:
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号