Control of the Total Artificial Heart: New Aspects in Human versus Animal Experience |
| |
Authors: | Heinrich Schima,Wolfgang Trubel,Farag Coraim,Leopold Huber,Michael Rolf Mü ller,Gerhard Redl,Udo Losert,Herwig Thoma,Ernst Wolner |
| |
Affiliation: | Second Department of Surgery, University of Vienna, Austria. |
| |
Abstract: | Control strategies for total artificial heart application have generally been based on experience with healthy animals. Human patients in a bad state of health who have impaired organ functions and who are subjected to intensive care procedures can develop atypical hemodynamic behavior. In these patients, both unstable and hyperstable behavior of the vascular resistance were observed. Therefore, regulation of cardiac output (CO) by pressure parameters only was avoided and CO was adjusted to obtain an appropriate O2-utilization (O2U). Intending to keep the O2U within ranges of 20-25%, we obtained cardiac indexes between 3.3 and 4.4 L/m2/min (CO 6-8 L/min), which is higher than other cardiac indexes reported. A CO of 10.5 L/min was even necessary to obtain an O2U of 30% in a septic patient. This strategy caused a stable driving management and led to a rapid hemodynamic stabilization and general improvement of the patients' condition. Results indicate that it is also very important to monitor metabolic parameters for appropriate driver adjustment as well, especially in the early postoperative phase, and that O2-U is a sensitive and useful parameter for this purpose. |
| |
Keywords: | TAH control Vasoregulation Postoperative condition Oxygen utilization O2X-ratio |
|
|