首页 | 本学科首页   官方微博 | 高级检索  
     


Late fiber metal shedding of the first and second-generation Harris Galante acetabular component. A report of 5 cases
Authors:Mayman David J  González Della Valle Alejandro  Lambert Edward  Anderson John  Wright Timothy  Nestor Bryan  Sculco Thomas P  Salvati Eduardo A
Affiliation:Hospital for Special Surgery, Weill Medical College, Cornell University, New York, New York 10021, USA.
Abstract:Five patients presented with fiber metal mesh shedding of a Harris Galante II acetabular cup detected between 11 and 15 years after implantation. All patients presented with hip pain and 4 demonstrated gross acetabular loosening and fiber metal separation on preoperative radiographs. The remaining patient underwent revision surgery because of a liner dislodgment and had a radiographically well fixed shell. Loosening and fiber metal separation were detected intraoperatively. Scanning electron microscopy of the retrieved shells demonstrated isolated diffusion bonding marks in the areas where the mesh separated from the substrate and no evidence of corrosion. Progressive osteolysis in the iliac bone was evident in 4 of our cases. Progressive iliac osteolysis may lead to loss of bone support in well-fixed cups and excessive stresses transferred to the interface between the fiber metal mesh and the titanium substrate leading to the separation of the 2 layers. Fiber metal separation may contribute to long-term loosening in the Harris Galante acetabular component.
Keywords:fiber metal separation   Harris Galante acetabular component   scanning electron microscopy   progressive osteolysis
本文献已被 ScienceDirect PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号