An Improved Method for Collecting Breath for the Assay of Acetaldehyde |
| |
Authors: | Michael Phillips MRCP Preston Hensley PhD Richard A Baiter MD Stanley L Cohan MD |
| |
Institution: | Departments of Medicine, Biochemistry, and Neurology;Georgetown University Medical Center, Washington, DC. |
| |
Abstract: | An improved method is described for the collection of breath for the subsequent assay of acetaldehyde and other volatile components. Breath is collected in a Pyrex gas-collecting tube sealed at both ends with Teflon taps. Prior to collection or assay of the samples, this tube is heated to 72 degrees C; breath is sampled for assay by piercing a rubber septum on a sideport with the needle of a similarly heated gas-tight syringe, and injected into a gas chromatograph (GC). The advantages of this system are: (1) Avoidance of the artefacts encountered in the assay of acetaldehyde in the blood; (2) suitability for sample collection at a site remote from the GC laboratory; (3) avoidance of sample loss by leakage, contamination, or partitioning into water condensed from breath; and (4) compatibility with a "nondedicated" GC lacking any special gas-collecting circuitry. A typical study of a normal human volunteer is described, demonstrating the rise and fall of the concentration of acetaldehyde and ethanol in the breath following the ingestion of an oral dose of ethanol. |
| |