首页 | 本学科首页   官方微博 | 高级检索  
检索        


Flexor tendon repair using a stainless steel external splint. Biomechanical study on human cadaver flexor tendons
Authors:Gordon L  Dysarz F A  Venkateswara K T  Mok A P  Ritchie R O  Rabinowitz S
Institution:Department of Orthopedic Surgery, California Medical Center, San Francisco, USA. lghand@aol.com
Abstract:A stainless steel external tendon splint was used in repair of cadaver tendons and compared with standard tendon repairs with suture. The splint was combined with a Kessler repair and tested against the Kessler, Becker, and Savage repairs in fresh human cadaver flexor digitorum profundus tendons. Biomechanical testing was done on a tensile testing machine, and load-displacement curves were generated. The repairs using the external tendon splint demonstrated a range of improvement of 32 to 146% in mean maximal tensile strength and a 20 to 185% improvement of mean ultimate tensile strength compared with all other repairs. The external tendon splint is relatively easy to apply to a tendon. The repair is strengthened and becomes capable of withstanding early active range of motion exercises. In vivo testing will be needed to assess the potential clinical usefulness of such a device.
Keywords:
本文献已被 ScienceDirect PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号