Dislodged and mechanically distorted stent stuck within a previously implanted drug-eluting stent
Authors:Hiroaki Watabe  Akira Sato  Tomoya Hoshi  Kenichi Obara  Ryo Kawamura  Kazutaka Aonuma
Institution:1. Cardiovascular Division, Faculty of Medicine, University of Tsukuba, 1-1-1 Tennodai, Tsukuba, Ibaraki, 305-8575, Japan
2. Department of Cardiology, Ryugasaki Saiseikai Hospital, Ryugasaki, Japan
Abstract:Coronary stent loss during percutaneous coronary intervention is rare and is often associated with significant morbidity. Several retrieval techniques, overlying stent deployment and crushing, and surgical removal can be used to deal with a stent lost in the coronary system. We successfully treated a dislodged and mechanically distorted coil stent stuck within a previously implanted drug-eluting stent (DES) by stent-crush technique. This case might provide insight into the mechanisms responsible for the longitudinal fragility of cobalt alloy and coil-structure stents and stent fracture of DES. In the DES era, careful attention should be paid to such complications when attempting to deliver a stent to a distal vessel through a pre-existing DES.
Keywords:
本文献已被 SpringerLink 等数据库收录!