首页 | 本学科首页   官方微博 | 高级检索  
     


Comparison of assays for determination of peptide content for lyophilized thymalfasin*
Authors:S. Vemuri
Abstract:Abstract: Precise determination of the peptide content in drug substance samples depends highly upon the particular peptide compound and methodology used. Four independent methods were evaluated and compared to determine which would produce the best experimental precision for analysis of thymalfasin (thymosin α‐1). Four different methods were evaluated including elemental analysis (CHN), quantitative amino acid analysis (AAA), high‐performance liquid chromatography (HPLC), and Kjeldahl. This study demonstrates that the AAA method is highly variable in one laboratory while quite precise in another laboratory. Similarly, HPLC results depended on the laboratory conducting the study with more precise values obtained under cGMP. On the contrary, the CHN method yielded highly precise [i.e. <2% coefficient of variation (CV)] values. As precise knowledge of protein content is fundamental for the compounding of final pharmaceutical product of a specific potency, the CHN analysis is recommended for peptide content determination of the drug substance thymalfasin.
Keywords:amino acid analysis  elemental analysis  Kjeldahl nitrogen  peptide content  precision  reverse phase high‐performance liquid chromatography  thymalfasin