PET/CT图像基于CT衰减校正伪影的实验研究
引用本文:陈泽龙,谢康宁,漆家学,等.PET/CT图像基于CT衰减校正伪影的实验研究[J].医疗卫生装备,2014(3):81-84.
作者姓名:陈泽龙  谢康宁  漆家学  
作者单位:[1]第四军医大学生物医学工程系,西安710032 [2]南京军区福州总医院医学影像中心,福州350025
基金项目:全军后勤科研计划项目面上课题资助(CWS11C148)
摘    要:目的:研究不同浓度泛影葡胺溶液在不同CT扫描条件下对PET/CT图像质量的影响。方法:向5个乳胶容器中分别注入体积分数为5%、10%、15%、20%和40%的1:7服泛影葡胺溶液,依次将其密闭置于圆柱形水模内进行PET/CT成像。利用CT扫描数据对PET图像进行衰减校正,并比较不同管电压CT扫描条件下PET图像的差异。结果:体积分数为15%、20%和40%的泛影葡胺充盈区在经CT衰减校正的PET图像上表现为高浓聚伪影.标准摄取值比本底均有不同程度的升高,而体积分数为5%、10%的泛影葡胺充盈区则无明显伪影,反映了随着泛影葡胺浓度的升高CT衰减校正伪影随之增强;对于体积分数为40%的口服泛影葡胺溶液,随着CT扫描所使用的管电压的升高.经CT衰减校正的PET图像上的伪影随之减弱.标准摄取值降低。结论:体积分数大于15%的泛影葡胺溶液在PET/CT成像中标准摄取值升高,图像表现出伪影;PET/CT成像中CT扫描所采用的管电压越高,高浓度泛影葡胺溶液产生的伪影越弱。

关 键 词:PET  CT  伪影  标准摄取值  图像质量

PET/CT Images Based on CT Attenuation Correction Artifact
CHEN Ze-long,XIE Kang-ning,QI Jia-xue,QIAN Gen-nian,ZHAO Chun-lei,CHEN Jie.PET/CT Images Based on CT Attenuation Correction Artifact[J].Chinese Medical Equipment Journal,2014(3):81-84.
Authors:CHEN Ze-long  XIE Kang-ning  QI Jia-xue  QIAN Gen-nian  ZHAO Chun-lei  CHEN Jie
Institution:2 (1.School of Biomedical Engineering, the Fourth Military Medical University, Xi'an 710032, China; 2.Department of Medical Imaging, Fuzhou General Hospital of Nanjing Military Area Command, Fuzhou 350025, China)
Abstract:Objective To study the influence of different concentrations of meglumine solutions on PET/CT image quality with different CT protocols. Methods Meglumine solutions with concentrations of 5%, 10%, 15%, 20% and 40% were injected into five latex vessels, then were sealed into a cylindrical water phantom. The phantom was scanned by PET/CT with different CT tube voltages, followed by a conventional attenuation correction procedure. Comparison was finally made among the images of different tube voltages. ~ There were artifacts in the areas filled with meglumine solutions with the concentrations Of 15%, 20% and 40%. Standard Uptake Value (SUV) grew higher than the background with varying degrees. However, there were no apparent ar- tifacts when the concentration of the meglumine solution was 5% or 10%. It's could be concluded that CT attenuation correction artifact became apparent with the increased concentration of meglumine solutions. For the meglumine solution with the concen- tration of 40%, as CT voltage rose the CT attenuation correction artifacts on PET images became much less apparent, and the SUV decreased. Conclusion The SUV on the PET/CT images of meglumine solutions with concentrations of more than 15% is higher than the background, and there are artifacts on the images. The artifact for high-concentration meglumine solution gets less apparent with the increase of the CT voltage, when PET/CT scanning performed.
Keywords:PET/CT  artifact  standard uptake value  image quality