Dual-energy CT for the evaluation of silicone breast implants
Authors:Thorsten R C Johnson  Isabelle Himsl  Karin Hellerhoff  Doris Mayr  Dorothea Rjosk-Dendorfer  Nina Ditsch  Bernhard Krauss  Klaus Friese  Maximilian F Reiser  Miriam S Lenhard
Institution:1. Department of Clinical Radiology, University of Munich-Grosshadern Campus, Marchioninistr. 15, 81377, Munich, Germany
2. Department of Obstetrics and Gynecology, University of Munich, Munich, Germany
3. Department of Pathology, University of Munich, Munich, Germany
4. Siemens AG, Healthcare Sector, Erlangen, Germany
Abstract:

Objective

The evaluation of breast implants for rupture is currently the domain of ultrasound and MRI, while mammography is of very limited diagnostic value. Recently, specific visualisation of silicone has become feasible using dual-energy CT. Our objective was to evaluate whether it is feasible to identify silicone in breast implants by dual-energy CT and to reliably diagnose or rule out ruptures.

Methods

Seven silicone breast implant specimens were examined on dual-source CT at 100- and 140-kV tube potential with a 0.8-mm tin filter (collimation 128 × 0.6 mm, current–time products 165 and 140 mAsref with modulation, rotation time 0.28 s, pitch 0.55). Two patients scheduled for implant removal or replacement were examined with identical parameters.

Results

The silicone of the implant specimens showed a strong dual-energy signal. In one patient, both implants were intact, while a rupture was identified in the other patient. Ultrasound, MRI, surgical findings and histology confirmed the dual-energy CT diagnosis.

Conclusion

Dual-energy CT may serve as an alternative technique for speedy evaluation of silicone breast implants. Specific clinical studies are required to determine the diagnostic accuracy and define indications for this technique.

Key Points

? Dual-energy CT makes it possible to visualise silicone in breast implants. ? Silicone provides a strong photoelectric effect that can be detected. ? Initial experience suggests that implant ruptures can be identified or ruled out.