Feasibility and reliability of DEXA analysis after total ankle arthroplasty: A cadaver study |
| |
Authors: | Elina Venjakob Daiwei Yao Sarah Ettinger Leif Claassen Michael Schwarze Christian Plaass Christina Stukenborg-Colsman Matthias Lerch |
| |
Affiliation: | 1. Department for Orthopedic Surgery, Hannover Medical School, Anna-von-Borries-Str. 1-7, 30625, Hannover, Germany;2. Laboratory for Biomechanics and Biomaterials, Department of Orthopaedics, Hannover Medical School, Anna-von-Borries-Str. 1-7, 30625 Hannover, Germany |
| |
Abstract: | BackgroundAlthough the outcomes of total ankle arthroplasty (TAA) have improved, unsolved problems such as stress shielding remain. Although dual energy X-ray absorptiometry (DEXA) is the “gold standard” for evaluation of these issues, it is rarely used in patients after TAA. This study aimed to establish a scan technique and to assess the mechanical changes in bone density caused by bone stock preparation.MethodsEight fresh-frozen cadaver legs were investigated by DEXA before TAA, with implant in situ, and after implant removal. Scan surface, bone mineral content, and bone mineral density were analysed to assess mechanical bone mass changes.ResultsWe examined data for density changes by bone compression after TAA, and found “good” results for internal reliability but only “acceptable” results for external reliability.ConclusionsThe results were reliable and reproducible. Using the present data, mechanical and biological processes can be considered together to understand the postoperative phases of bone remodelling after TAA. |
| |
Keywords: | Total ankle arthroplasty (TAA) Dual energy X-ray absorptiometry (DEXA) Stress shielding Bone remodelling |