大连市2017年医疗机构医用辐射防护监测结果分析
引用本文:张兴晖,倪长玉,董倩倩,叶倩鸥,刘双.大连市2017年医疗机构医用辐射防护监测结果分析[J].中国辐射卫生,2019,28(4):414-416.
作者姓名:张兴晖  倪长玉  董倩倩  叶倩鸥  刘双
作者单位:大连市疾病预防控制中心, 辽宁 大连 116021
摘    要:目的 了解大连市2017年医疗机构(除口腔诊所或门诊)开展放射诊疗服务情况、接受X射线诊断的患者或受检者情况和监测点医疗机构的辐射防护情况。方法 通过对大连市所有医疗机构的调查和对15家监测点医疗机构的监测,对全市医疗机构开展放射诊疗服务、接受X射线诊断的患者或受检者以及监测点医疗机构的放射工作场所辐射防护检测、X射线装置质量控制检测情况进行分析评价。结果 结果显示全市277家医疗机构均开展了X射线诊断学,开展率为100%;开展介入放射学的有14家,占5.1%;开展核医学的有5家,占1.8%;开展放射治疗的有8家,占2.9%。全市医疗机构使用X射线诊断的人次数为4 140 427,其中常规X射线诊断2 285 306人次,占55.2%;CT诊断1 740 424人次,占42.0%;其它诊断114 697人次,占2.8%。15个监测点医疗机构、141个放射工作场所辐射防护水平均合格,X射线装置的性能检测均合格。结论 CT机的使用率非常高,应注意CT机使用的正当性,并加强辐射防护的管理。

关 键 词:医疗机构  辐射防护  监测  分析  
收稿时间:2019-01-07

Analysis of medical radiation protection monitoring results of medical institutions in Dalian in 2017
ZHANG Xinghui,NI Changyu,DONG Qianqian,YE Qianou,LIU Shuang.Analysis of medical radiation protection monitoring results of medical institutions in Dalian in 2017[J].Chinese Journal of Radiological Health,2019,28(4):414-416.
Authors:ZHANG Xinghui  NI Changyu  DONG Qianqian  YE Qianou  LIU Shuang
Institution:Dalian Center for Disease Control and Prevention, Dalian 116021 China
Abstract:Objective To understand the medical institutions (except oral clinics or outpatient clinics) in dalian city in 2017 to carry out radiological diagnosis and treatment services, the patients or examinees receiving X-ray diagnosis and the radiation protection situation of the medical institutions at the monitoring points.Methods Through the survey of dalian all medical institutions and to 15 medical institutions of monitoring, to the city of radiological diagnosis and treatment of medical institutions and service, or subjects and in patients undergoing X-ray diagnostic monitoring radioactive workplace radiation protection of medical institutions detection, X-ray equipment quality control testing situation analysis evaluation.Results The results showed that all 277 medical institutions in the city had carried out X-ray diagnosis, with the development rate of 100%. Fourteen medical institutions had carried out interventional radiology, accounting for 5.1%. Five medical institutions, or 1.8 percent, had carried out nuclear medicine; Eight medical institutions, or 2.9%, had carried out radiation therapy. The number of times using X-ray diagnosis in medical institutions in the city was 4 140 427, among which 2 285 306 person times received conventional X-ray diagnosis, accounting for 55.2%. CT diagnosis reached 1 740 424 person-times, accounting for 42.0%. Other diagnoses were 114 697, or 2.8%. The radiation protection level of 15 medical institutions and 141 radiation workplaces at the monitoring sites were all qualified, and the performance test of the X-ray device was all qualified.Conclusion The utilization rate of CT machines is very high, so we should pay attention to the legitimacy of the use of CT machines and strengthen the management of radiation protection.
Keywords:Medical Institutions  Radiation Protection  Monitoring  Analysis  