首页 | 本学科首页   官方微博 | 高级检索  
     

两种CT胸部扫描模式下DLP值的测量与分析
引用本文:刘芳,王小山,陈英民,宋钢,刘乾,牛菲. 两种CT胸部扫描模式下DLP值的测量与分析[J]. 中国辐射卫生, 2019, 28(6): 633-636. DOI: 10.13491/j.issn.1004-714x.2019.06.010
作者姓名:刘芳  王小山  陈英民  宋钢  刘乾  牛菲
作者单位:山东省医学科学院放射医学研究所 山东第一医科大学, 山东 济南 250062
基金项目:山东省医药卫生科技发展计划项目(2016WSA18040);山东省医学科学院医药卫生创新工程
摘    要:目的 通过对CT胸部常规扫描与低剂量扫描模式下的剂量长度乘积(DLP)值进行比较,为CT胸部扫描时的条件选择提供数据参考。方法 采用CT-DP检测探头和标准CT剂量体模,测量GE Revolution型256排CT胸部常规扫描模式与低剂量扫描模式下的DLP。结果 对胸部低剂量模式,Chest_Lowdose LARGE SIZE模式,DLP测量值60.16 mGy·cm;Chest_Lowdose SMALL SIZE模式,DLP测量值33.65 mGy·cm。胸部常规扫描模式,DLP测量值276.68 mGy·cm。胸部常规扫描模式DLP值是Chest_Lowdose SMALL SIZE模式的8.22倍,是Chest_Lowdose LARGE SIZE模式的4.60倍。而同为胸部低剂量扫描,根据受检者体型不同,选择Chest_Lowdose LARGE SIZE模式DLP值是Chest_Lowdose SMALL SIZE模式的1.79倍。结论 选择低剂量CT进行胸部扫描能显著降低受检者剂量,对体型瘦小的受检者选择合适的胸部低剂量扫描模式也会明显降低所受剂量。

关 键 词:CT剂量分析仪  剂量长度乘积  低剂量CT扫描  
收稿时间:2019-05-07

Measurement and analysis of DLP values in two chest CT scanning modes
LIU Fang,WANG Xiaoshan,CHEN Yingmin,SONG Gang,LIU Qian,NIU Fei. Measurement and analysis of DLP values in two chest CT scanning modes[J]. Chinese Journal of Radiological Health, 2019, 28(6): 633-636. DOI: 10.13491/j.issn.1004-714x.2019.06.010
Authors:LIU Fang  WANG Xiaoshan  CHEN Yingmin  SONG Gang  LIU Qian  NIU Fei
Affiliation:Institute of Radiation Medicine, Shandong Academy of Medical Sciences Shandong First Medical University, Jinan 250062 China
Abstract:Objective By comparing the dose length product (DLP) values of routinechest CT scanand low-dose CT chest scan, we can provide data for the selection of chest CT scan conditions.Methods Using a CT Dose Profiler(CT-DP) and a standard CT dose phantom, we measured DLPs in the routineand low-dose chest CT scanning modes by GE revolution 256-slice CT.Results Forlow-dose chest scanning modes:in Chest_Lowdose LARGE SIZE mode, DLP measured was 60.16 mGy·cm; in Chest_Lowdose SMALL SIZE mode,DLP was 33.65 mGy·cm;and for routine chest scan, DLP was 276.68 mGy·cm.The DLP value ofthe routine chest scanning mode is 8.22 times larger than that ofChest_Lowdose SMALL SIZE mode, and 4.60 times larger than that in the Chest_Lowdose LARGE SIZE mode.And for the two low-dose chest scans, depending on the size of the subject,the DPL of Chest_Lowdose LARGE SIZE mode was 1.79 times larger than that of Chest_Lowdose SMALL SIZEmode.Conclusion Low-dose chest scan can significantly reduce the dose of the subject, and choosing a suitable low-dose chest scan for the subject with a small size will also significantly reduce the dose.
Keywords:CT Dose Profiler(CT-DP)  Dose Length Product(DLP)  Low-Dose CT Scan  
本文献已被 CNKI 等数据库收录!
点击此处可从《中国辐射卫生》浏览原始摘要信息
点击此处可从《中国辐射卫生》下载全文
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号