Validity of a hinged constant force probe and a similar, immobilised probe in untreated periodontal disease |
| |
Authors: | Paul Simons Trevor Watts |
| |
Affiliation: | Department of Periodontology, U.M.D.S., London, UK. |
| |
Abstract: | The validity of a hinged constant force probe (0.25 N) was compared with that of a similar but immobilised instrument, using the same interchangeable tip for both (0.64 mm diameter; 2 mm divisions). 60 sites were measured on teeth which were extracted subsequently, in patients with untreated periodontal disease, and the connective tissue attachment level was used as validity criterion. The clinical measurements of both probes correlated well with each other, but they differed significantly from the post-extraction connective tissue attachment level measurements, indicating a point 1.2 mm coronally to this, on average. A companion investigation of intra-operator probing depth reproducibility with the 2 probes, was undertaken in 14 patients, at 2 visits separated by 1 week in each case. All patients had untreated periodontal disease. A difference between probes was found at the first visit, but not at the second; the immobilised probe showed a difference between visits, reducing mean probing depth slightly at the second; when the immobilised probe was used first, there was a difference between probes. Further analysis of the results indicated that there was greatest agreement between probes when the constant force probe had been used before the immobilised probe at the second visit. The results suggested that these probes indicated a point above the connective tissue attachment level, related to pocket morphology, and that there was a moderate learning effect due to operator use of the constant force probe, which modified use of the immobilised probe. |
| |
Keywords: | constant force probe validity learning effect intra-operator reproducibility |
|
|