Comparison of anterior segment measurements using rotating Scheimpflug imaging and partial coherence interferometry
Authors:Mohammad Soleimani  Hassan Hashemi  Shiva Mehravaran  Mehdi Khabazkhoo  Mohammad Hassan Emamian  Mohammad Shariati  Akbar Fotouhi
Affiliation:Rasoul-e-Akram Hospital, Tehran University of Medical Sciences, Tehran, Iran;Noor Ophthalmology Research Center, Noor Eye Hospital, Tehran, Iran;Noor Ophthalmology Research Center, Noor Eye Hospital, Tehran, Iran;Noor Ophthalmology Research Center, Noor Eye Hospital, Tehran, Iran;Shahroud University of Medical Sciences, Shahroud, Iran;Education Development Center, Tehran University of Medical Sciences, Tehran, Iran;Department of Epidemiology and Biostatistics, School of Public Health, Tehran University of Medical Sciences, Tehran, Iran
Abstract:

AIM

To compare central corneal thickness (CCT) and anterior chamber depth (ACD) measurements using rotating Scheimpflug imaging and partial coherence interferometry.

METHODS

As part of the first phase of Shahroud Eye Cohort Study with 5 190 subjects of 40 to 64 years of age, CCT and ACD were measured using Scheimpflug imaging with the Pentacam (Oculus, Inc., Lynnwood, WA, USA) and partial coherence interferometry with the Allegro BioGraph (Wavelight, Erlangen, Germany).

RESULTS

After applying exclusion criteria, we had data of 4 387 subjects with a mean age of 50.7±6.2 years. Mean CCT with Pentacam and BioGraph were 528.6±33.2µm and 525.6±32µm respectively; the difference was statistically significant (P<0.001), but the correlation was high (R=0.920). Mean ACD measurements using Pentacam and BioGraph were 2.68±0.35mm and 2.62±0.33mm respectively; the inter-device difference was significant (P<0.001) with high correlation (R=0.944). The 95% limits of agreements between devices were -22.65µm to 28.61µm and -0.16mm to 0.29mm for CCT and ACD measurements, respectively.

CONCLUSION

For both CCT and ACD, the BioGraph gave significantly lower values than the Pentacam (P<0.05). Despite the high inter-device correlation, the 95% limits of agreements were wide, and this may limit their interchangeability in measuring the CCT and ACD.
Keywords:central corneal thickness   anterior chamber depth   Pentacam, BioGraph   agreement