首页 | 本学科首页   官方微博 | 高级检索  
     


VALIDATION OF A SYSTEM FOR MEASUREMENT OF METABOLIC GAS EXCHANGE DURING ANAESTHESIA WITH CONTROLLED VENTILATION IN AN OXYGEN CONSUMING LUNG MODEL
Authors:SVENSSON, K. L.   SONANDER, H. G.   STENQVIST, O.
Affiliation:Department of Anaesthesia and Intensive Care, Sahlgren's Hospital, University of G"o"teborg S-413 45 Göteborg, Sweden
Abstract:An oxygen consuming lung model was used for evaluation and validationof a technique for metabolic gas exchange measurements duringcontrolled ventilation. The technique comprised a Servo 900C ventilator (Siemens) and separate oxygen and carbon dioxideanalysers (Beck-man). Measurements of oxygen consumption andcarbon dioxide production were made either by measuring inspiredand expired ventilation and gas fractions in these volumes orby measuring gas fractions and calculating expired ventilationfrom inspired by transformation (Haldane). Irrespective of theFlO2, measured values correlated well with lung model settings:measured values were within ±2% of simulated. When Haldanetransformation was used with an FlO2 of 0.5 there was a significantunderestimation of oxygen consumption. Carbon dioxide productionvalues correlated well irrespective of the FlO2 used or methodof measurement of ventilation volume. Metabolic gas exchangemeasurements by measuring both inspired and expired ventilationvolumes may be used when inert gases are not in equilibrium,for example during nitrous oxide anaesthesia.
Keywords:
本文献已被 Oxford 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号