GE Signa Twinspeed 1.5T磁共振肝脏扩散加权成像的技术探讨
引用本文:尚全良,肖恩华,贺忠,谭利华,司徒卫军,白末了,袁术文,杜万平. GE Signa Twinspeed 1.5T磁共振肝脏扩散加权成像的技术探讨[J]. 中国医学影像技术, 2004, 20(10): 1515-1518
作者姓名:尚全良  肖恩华  贺忠  谭利华  司徒卫军  白末了  袁术文  杜万平
作者单位:中南大学湘雅二医院放射科,湖南,长沙,410011
基金项目:国家自然科学基金,湖南省中医药管理局科研项目
摘    要:目的探讨高场强双梯度磁共振机(GE Signa Twinspeed 1.5T)的肝脏扩散加权成像技术(DWI).方法对20例正常肝脏进行扩散加权成像研究,分别改变重复时间(TR)及扩散敏感系数(b值),测定ADC值、信噪比、图像质量等.结果 b值固定,TR值改变,正常肝组织平均ADC值差异不显著(P>0.05);TR值固定,b值改变,ADC值随b值增大而减小(P<0.05,差异有显著性).b值越大,图像质量越差;b≥800 s/mm2时,图像几乎难以观察.结论 ADC值随b值的增大而减小.行肝脏扩散加权成像时,可适当减小TR值,以缩短扫描时间及患者的屏气时间,减少呼吸运动伪影.

关 键 词:肝脏  磁共振成像  扩散
文章编号:1003-3289(2004)10-1515-04
修稿时间:2004-04-07

Study of technology of liver diffusion-weighted imaging at GE Signa Twinspeed 1.5
T MR SHANG Quan-liang,XIAO En-hu,HE zhong,TAN Li-hu,SITU Wei-jun,BAI Mo-liao,YUAN Shu-wen and DU Wan-ping. Study of technology of liver diffusion-weighted imaging at GE Signa Twinspeed 1.5[J]. Chinese Journal of Medical Imaging Technology, 2004, 20(10): 1515-1518
Authors:T MR SHANG Quan-liang  XIAO En-hu  HE zhong  TAN Li-hu  SITU Wei-jun  BAI Mo-liao  YUAN Shu-wen  DU Wan-ping
Affiliation:Department of Radiology, the Second Xiangya Hospital, Central South University, Changsha 410011, China;Department of Radiology, the Second Xiangya Hospital, Central South University, Changsha 410011, China;Department of Radiology, the Second Xiangya Hospital, Central South University, Changsha 410011, China;Department of Radiology, the Second Xiangya Hospital, Central South University, Changsha 410011, China;Department of Radiology, the Second Xiangya Hospital, Central South University, Changsha 410011, China;Department of Radiology, the Second Xiangya Hospital, Central South University, Changsha 410011, China;Department of Radiology, the Second Xiangya Hospital, Central South University, Changsha 410011, China;Department of Radiology, the Second Xiangya Hospital, Central South University, Changsha 410011, China
Abstract:Objective To study the technology of liver diffusion-weighted imaging (DWI) at 1.5T MR (GE Signa Twinspeed). Methods DWI and measurement of apparent diffusion coeffecient (ADC) value, signal-to-noise ratio (SNR) and the quality of images were performed in 20 healthy subjects by changing TR value and b value at 1.5T MRI. Results When b value was not changed and TR value was changed, the ADCs of normal livers were insignificantly different ( P >0.05), but when TR value was not changed and b value was changed, the ADCs of normal livers were significantly different ( P < 0.05 ). The ADC decreased as the b value increased. The higher the b value became, the worse the quality of the DWI. When b value equal or exceed 800 s/mm 2, the quality of the DWI was too bad to be analyzed. Conclusion ADC value decreases as b value increases. To reduce the time of holding breath and artifacts of the breath and to improve the accuracy of the ADC value and the quality of DWI, we suggest that TR value can be appropriately reduced in liver DWI.
Keywords:Liver  Magnetic resonance imaging  Diffusion