A Device and a Method for Simulating Supraventricular Arrhythmias in Pacemakers Implanted in Patients: Assessment of Safety and Reliability
Authors:Francesca Tronconi  Maria Cristina Porciani  Lorella Gianfranchi  Giorgio Corbucci  Luigi Padeletti
Affiliation:(1) Institute of Internal Medicine and Cardiology, University of Florence, Florence, Italy;(2) Department of Cardiology, General Hospital, Cento (FE), Italy;(3) Vitatron Medical Italia, Bologna, Italy;(4) Institute of Internal Medicine and Cardiology, University of Florence, Florence, Italy
Abstract:Automatic mode switching, enabling the pacemaker to pace at an independent ventricular rate when atrial fibrillation occurs, was introduced to provide protection against rapid ventricular pacing during atrial arrhythmia. This study proposes a technique to test automatic mode switching performance. It is based on a programmable system (Arrhythmia Simulator) generating pulse trains that, when applied to the patient's skin, may interfere with the implanted device simulating supraventricular arrhythmias. The amplitude of the output signals is 5 V and they are delivered through an output resistance of 50 kOHgr limiting the current to 100 mgrA that guarantees no risk of muscle stimulation during the test and meanwhile low noise signals on the surface ECG. The duration of pulses is 20 ms. Pulses delivered by the Arrhythmia Simulator were correctly sensed by the pacemaker's atrial channel, and the detected amplitude was 2 ± 0.8 mV, ranging from 1 to 3 mV. The performance of the system was reliable and safe in every patient despite the use of different pacemaker leads. Automatic mode switching was induced in every patient at every attempt. The pulses delivered by Arrhythmia Simulator didn't alter the quality of ECG tracings making easy and reliable the analysis.
Keywords:simulation of supraventricular arrhythmias  atrial fibrillation  pacemaker
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号