Comparing calibration methods of electron beams using plane-parallel chambers with absorbed-dose to water based protocols
Authors:Stewart K J  Seuntjens J P
Institution:Department of Medical Physics, McGill University Health Centre, Montreal General Hospital, Canada. kristins@medphys.mcgill.ca
Abstract:Recent absorbed-dose-based protocols allow for two methods of calibrating electron beams using plane-parallel chambers, one using the N(Co)D,w for a plane-parallel chamber, and the other relying on cross-calibration of the plane-parallel chamber in a high-energy electron beam against a cylindrical chamber which has an N(Co)D,w factor. The second method is recommended as it avoids problems associated with the Pwall correction factors at 60Co for plane-parallel chambers which are used in the determination of the beam quality conversion factors. In this article we investigate the consistency of these two methods for the PTW Roos, Scanditronics NACP02, and PTW Markus chambers. We processed our data using both the AAPM TG-51 and the IAEA TRS-398 protocols. Wall correction factors in 60Co beams and absorbed-dose beam quality conversion factors for 20 MeV electrons were derived for these chambers by cross-calibration against a cylindrical ionization chamber. Systematic differences of up to 1.6% were found between our values of Pwall and those from the Monte Carlo calculations underlying AAPM TG-51, and up to 0.6% when comparing with the IAEA TRS-398 protocol. The differences in Pwall translate directly into differences in the beam quality conversion factors in the respective protocols. The relatively large spread in the experimental data of Pwall, and consequently the absorbed-dose beam quality conversion factor, confirms the importance of the cross-calibration technique when using plane-parallel chambers for calibrating clinical electron beams. We confirmed that for well-guarded plane-parallel chambers, the fluence perturbation correction factor at d(max) is not significantly different from the value at d(ref). For the PTW Markus chamber the variation in the latter factor is consistent with published fits relating it to average energy at depth.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号