CT灌注测量中去卷积算法准确性的检验
引用本文:袁小东,张静,田建明,陈炜,徐志江.CT灌注测量中去卷积算法准确性的检验[J].中国医学影像技术,2005,21(5):802-805.
作者姓名:袁小东  张静  田建明  陈炜  徐志江
作者单位:1. 第二军医大学长海医院放射科,上海,200433
2. 常州市武进县人民医院放射科,江苏常州,213000
摘    要:目的验证去卷积算法在组织灌注测量中所得结果的准确性.方法 30例健康成人行左肾静脉水平肾皮质CT灌注扫描,所得数据用去卷积算法计算灌注参数,由所得参数进一步推算肾静脉增强TDC,并与实际测得肾静脉TDC进行比较,计算决定系数R2,评价两者的吻合程度.统计工作在MS Excel 2000中完成.结果决定系数R2的均数的95%可信区间为0.85~0.91.理论计算所得肾静脉TDC与实测TDC符合程度较高.结论 CT组织灌注测量中去卷积算法是一种较准确的计算方法.

关 键 词:去卷积算法  灌注  体层摄影术  X线计算机  肾皮质
文章编号:1003-3289(2005)05-0802-04
收稿时间:2004/10/12 0:00:00
修稿时间:2004年10月12

Deconvolution arithmetic in CT perfusion measurement of renal cortex perfusion
YUAN Xiao-dong,ZHANG Jing,TIAN Jian-ming,CHEN Wei and XU Zhi-jiang.Deconvolution arithmetic in CT perfusion measurement of renal cortex perfusion[J].Chinese Journal of Medical Imaging Technology,2005,21(5):802-805.
Authors:YUAN Xiao-dong  ZHANG Jing  TIAN Jian-ming  CHEN Wei and XU Zhi-jiang
Institution:Department of Radiology, Changhai Hospital, Second Military Medical University,Shanghai 200433, China;Department of Radiology, Changhai Hospital, Second Military Medical University,Shanghai 200433, China;Department of Radiology, Changhai Hospital, Second Military Medical University,Shanghai 200433, China;Department of Radiology, Changhai Hospital, Second Military Medical University,Shanghai 200433, China;Department of Radiology,Wujin People's Hospital,Changzhou 213000, China
Abstract:Objective To validate the accuracy of results drawn by deconvolution arithmetic in CT perfusion measure. Methods Thirty healthy adults randomly collected underwent CT perfusion scan to acquire the perfusion data of the renal cortex at the level of left renal vein. Then deconvolution arithmetic was used to calculate perfusion parameters. According to the parameters, the enhanced TDC of renal vein was deduced and compared with actual TDC of renal vein. The coefficient of determination ( R 2 ) was calculated to evaluate the agreement degree of theoretical and actual TDC. Results The agreement degree between theoretical and actual TDC of renal vein was high, the 95% confidence interval of the mean of R 2 was 0.85-0.91 . Conclusion Deconvolution arithmetic is an accurate arithmetic method in tissue perfusin measured with CT.
Keywords:Deconvolution  Perfusion  Tomography  X-ray computed  Renal cortex