首页 | 本学科首页   官方微博 | 高级检索  
     


Ventricular Rate Stabilization Algorithm of ICD Causing Dual Chamber Pacing During Ventricular Tachycardia
Authors:S. Serge Barold
Affiliation:(1) Tampa General Hospital, Tampa, FL, USA
Abstract:Some implantable cardioverter-defibrillators (ICD) are designed with a ventricular rate stabilization (VRS) algorithm triggered by sensing a pacemaker-defined ventricular premature complex, for the prevention of ldquoshort-long-shortrdquo sequences that may predispose to ventricular tachycardia (VT). The VRS algorithm initiates AV sequential pacing at a relatively fast and decreasing rate according to a programmable sequence. This report describes a complex manifestation of ICD function where activation of the VRS algorithm produced AV sequential pacing during sustained VT. This response occurs when there is an appropriate relationship between the timing cycles of the ICD and a relatively slow VT.
Keywords:cardioverter defibrillator  ventricular tachycardia  dual chamber pacing  ventricular rate stabilization
本文献已被 SpringerLink 等数据库收录!