首页 | 本学科首页   官方微博 | 高级检索  
     


Comparison of the thermal tissue effects produced by aged sapphire and silica hemispherical tips
Authors:David D. Royston  Jorge H. Torres  Sharon Thomsen  P. S. Sriram  Ashley J. Welch
Abstract:This study evaluated the performance of sapphire and fused silica hemispherical tips under the same exposure conditions. Lesions produced in the chicken breast and a blood field were sectioned for light and transmission polarizing microscopy. Lesion size and thermal damage area were recorded as a function of the tips accumulated exposure. The tips transmission was measured after every 1,000 J of exposure. Fused silica tips lasted for ~5,000 J and experienced significant surface and transmission deterioration. The sapphire hemispherical tips lasted for > 12,000 J with no surface and transmission deterioration. Lesions produced with the fused silica tips generally increased in depth with use, and depths of 6 mm were common. Lesions produced by the sapphire tips were subsurface spherical areas of coagulation with the tissue surface relatively intact. This difference in resulting lesions may be attributed to the higher thermal conductivity of sapphire. © 1994 Wiley-Liss, Inc.
  • 1 This article is a US Government work and, as such, is in the public domain in the United States of America.
  • Keywords:Nd:YAG laser  fiber optics  laser ablation  laser coagulation  chicken breast model  contact laser surgery
    设为首页 | 免责声明 | 关于勤云 | 加入收藏