首页 | 本学科首页   官方微博 | 高级检索  
     


Recording characteristics of the surface EMG electrodes
Authors:Paul E. Barkhaus  Sanjeev D. Nandedkar
Abstract:Routine motor nerve conduction studies are conducted using surface EMG electrodes. Most techniques of estimating the number of motor units (MUs) are based on surface EMG recordings. Therefore, it is important to assess the uptake area of these electrodes. We recorded surface EMG motor unit action potentials (SMUAPs) from the biceps muscle of normal subjects. The SMUAP amplitude fell from 42 μV for the superficially located MUs (i.e., within 10 mm of skin surface) to 11 μV for the deep MUs (i.e., more than 20 mm from the skin surface). We infer that the pickup radius of the surface electrode is less than 20 mm. The implications of the limited uptake area of the surface electrodes to the analysis of compound muscle action potentials, estimation of the number of MUs, and the surface EMG recordings are discussed. © 1994 John Wiley & Sons, Inc.
Keywords:surface EMG  EMG electrodes  uptake area  motor unit action potential  MU estimation
设为首页 | 免责声明 | 关于勤云 | 加入收藏