Variability of aortic valve calcification measurement with multislice spiral computed tomography |
| |
Authors: | Ruhl Karl M Das Marco Koos Ralf Mühlenbruch Georg Flohr Thomas G Wildberger Joachim E Günther Rolf W Mahnken Andreas H |
| |
Institution: | Department of Diagnostic Radiology, RWTH Aachen University, Germany. ruhl@rad.rwth-aachen.de |
| |
Abstract: | OBJECTIVES: We sought to assess the variability of aortic valve calcifications (AVCs) regarding the reconstruction window at different heart phases using multislice-spiral computed tomography. MATERIALS AND METHODS: A total of 46 patients (26 men; mean age. 65 years) underwent AVC scoring with multislice-spiral computed tomography (12 x 0.75 mm, 120 kV, 133 mAseff). Image reconstruction was performed every 10% of the RR-interval (0-90%). AVC was quantified using Agatston score, calcium volume, and calcium mass. Images were assessed for least motion artifacts. Coefficients of variation and Wilcoxon test were calculated. RESULTS: AVC scores are lowest at 60% and highest at 0% of the RR-interval (P < 0.001). Mean coefficients of variation were 36.2% (Agatston score), 38.7% (calcium volume), and 32.9% (calcium mass). At 60% (50-70%). minimal motion artifacts and the lowest variability of the scores were found. CONCLUSIONS: AVC scores show large variability depending on the point of image reconstruction. Diastolic image reconstruction at 60% of the RR-interval is recommended. |
| |