Simulated pulmonary nodules: detection with dual-energy digital versus conventional radiography |
| |
Authors: | Niklason, LT Hickey, NM Chakraborty, DP Sabbagh, EA Yester, MV Fraser, RG Barnes, GT |
| |
Abstract: | Performance of a prototype dual-energy digital chest radiography unit in detecting calcified and noncalcified simulated pulmonary nodules was compared with that of a highly optimized, conventional system. Nodules ranging in size (0.5, 1.0, and 1.6 cm), in number (five to 11), and in calcium content (0-25 mg) were superimposed over the lungs of a frozen, unembalmed, human chest phantom. For each technique, six observers examined 50 posteroanterior projections with different randomized nodule locations. Detection consisted of locating and assigning a level of confidence to each perceived nodular opacity. The resulting plots of the true-positive fraction versus the mean number of false-positive calls per projection indicate that for both calcified and noncalcified nodules, the digital unit performed significantly better (P less than .01). |
| |