Accelerated volumetric MRI with a SENSE/GRAPPA combination |
| |
Authors: | Blaimer Martin Breuer Felix A Seiberlich Nicole Mueller Matthias F Heidemann Robin M Jellus Vladimir Wiggins Graham Wald Lawrence L Griswold Mark A Jakob Peter M |
| |
Institution: | Department of Experimental Physics 5, University of Würzburg, Würzburg, Germany. martin.blaimer@case.edu |
| |
Abstract: | PURPOSE: To combine the specific advantages of the generalized autocalibrating partially parallel acquisitions (GRAPPA) technique and sensitivity encoding (SENSE) with two-dimensional (2D) undersampling. MATERIALS AND METHODS: By splitting the 2D reconstruction process into multiple one-dimensional (1D) reconstructions, the normal 1D GRAPPA method can be used for image reconstruction. Due to this data-handling process, a GRAPPA reconstruction is performed along the phase-encoding (PE) direction and effectively a SENSE reconstruction is performed along the partition-encoding (PAE) direction. RESULTS: In vivo experiments demonstrate the successful implementation of the SENSE/GRAPPA combination. Experimental results with up to 9.6-fold acceleration using a prototype 32-channel receiver head coil array are presented. CONCLUSION: The proposed SENSE/GRAPPA combination for 3D imaging allows the GRAPPA method to be applied in combination with 2D undersampling. Because the SENSE/GRAPPA combination is not based on knowledge of spatial coil sensitivities, it should be the method of choice whenever it is difficult to extract the sensitivity information. |
| |
Keywords: | 3D MRI phased array parallel imaging SENSE GRAPPA |
本文献已被 PubMed 等数据库收录! |