Evaluation of the Bio-Rad BioPlex Measles, Mumps, Rubella, and Varicella-Zoster Virus IgG multiplex bead immunoassay
Authors:Binnicker Matthew J  Jespersen Deborah J  Rollins Leonard O
Affiliation:Mayo Clinic, 200 First Street SW, SDSC 1-526, Rochester, MN 55905, USA. binnicker.matthew@mayo.edu
Abstract:The goal of this study was to compare the BioPlex 2200 measles, mumps, rubella, and varicella-zoster virus (MMRV) IgG multiplex assays (Bio-Rad Laboratories, Hercules, CA) to routine testing by enzyme immunoassay (EIA). Serum specimens (n = 500) submitted to our reference laboratory for routine MMRV IgG testing by EIA were also tested by the BioPlex assays. Following testing, the BioPlex measles, mumps, rubella, and varicella-zoster virus assays demonstrated agreements of 91.6% (95% confidence interval [CI], 88.8% to 93.7%), 94.2% (95% CI, 91.7% to 95.7%), 94.4% (95% CI, 92.0% to 96.1%), and 91.8% (95% CI, 89.0% to 93.9%), respectively, compared to the results of EIA. Timing studies showed that the BioPlex MMRV assay could provide complete analysis of 100 serum specimens in 1.7 h, compared to 5.5 h by EIA. These data indicate that the BioPlex MMRV IgG assays exhibit comparable performance (93% overall agreement [1,860/2,000 results]; κ = 0.67) to routine testing by EIA. The BioPlex assays allow for the simultaneous detection of all four analytes, thereby eliminating potential aliquot errors and reducing turnaround time.