首页 | 本学科首页   官方微博 | 高级检索  
     


Comparison of methods of attenuation and scatter correction in brain perfusion SPECT
Authors:Hayashi Masuo  Deguchi Jun  Utsunomiya Keita  Yamada Makoto  Komori Tsuyoshi  Takeuchi Masayasu  Kanna Kensei  Narabayashi Isamu
Affiliation:Department of Radiology, Osaka Medical College, Osaka, Japan. rad040@poh.osaka-med.ac.jp
Abstract:OBJECTIVE: The radioactivity count distribution in the brain must be determined accurately to accurately measure cerebral blood flow (CBF). Scatter and attenuation are factors that compromise the accuracy of determining radioactivity counts in the brain. METHODS: We compared regional CBF in patients by an autoradiographic method using N-isopropyl-p-[123I] iodoamphetamine when, first, attenuation correction alone was performed uniformly on SPECT images by using empiric mu-values (Chang method); second, scatter correction was performed and the mu-values of a homogeneous-attenuation body of water were used for attenuation correction (triple-energy window [TEW]+Chang method); and third, scatter correction was performed and the mu-values calculated by CT were used for attenuation correction (TEW+CT method). We also compared regional CBF measured by these methods with the values obtained by the xenon CT/CBF method, which uses CT and stable xenon. RESULTS: Scatter correction reduced overestimation of regional CBF in low-flow regions. The TEW+CT method yielded better regional and overall correlations with the xenon CT/CBF method than did either of the other methods. CONCLUSION: The TEW+CT method of correction gave the most accurate measurements of regional CBF.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号