Detection of cardiac-specific creatine kinase isoenzyme in sera with normal or slightly increased total creatine kinase activity.
Authors:D W Mercer  M A Varat
Abstract:We describe a spectrophotometric kinetic assay for detecting creatine kinase MB isoenzyme activity in the 1 to 10 U/liter range. The MB isoenzyme was isolated Clin. Chem. 20, 36 (1974)] and assayed (Rosalki method) with an Abbott ABA-100. Good reproducibility was demonstrated for MB isoenzyme activities near 1 U/liter (CV = 2.6%). Sera with normal or slightly increased total creatine kinase activity were evaluated. Sera of 14 patients with acute myocardial infarction contained, per liter, 84 to 236 U of total creatine kinase activity and 4.6 to 28.0 U of isoenzyme MB activity; corresponding ranges for sera from healthy lab technicians and patients with noncardiac disease were 36 to 277 and 0 to 2.6 U. MB isoenzyme activity for infarction patients rose and fell sharply within three days after the infarction. Atypical time-course patterns, MB isoenzyme activity remaining abnormally great for five days, were observed in serum from patients with prolonged atrial fibrillation and congestive heart failure or cardiomyopathy; the BB isoenzyme (1 to 5 U/liter) was also detected in sera of such patients but was absent in sera from infarcation patients. Quantification of column-isolated MB by the assay described is rapid, easy, specific, and extremely sensitive for measuring MB in the 1 to 10 U/liter range.