应用4650I型压缩机雾化器进行支气管激发试验的研究
引用本文:袁玉如,王佑娟.应用4650I型压缩机雾化器进行支气管激发试验的研究[J].华西医科大学学报,2000,31(2):262-264.
作者姓名:袁玉如  王佑娟
摘    要:为了解Models 4650I型压缩机雾化器(简称M)进行支气管激发试验的可行性,将92例支气管哮喘患者按用M雾化吸入时间的不同,分为1、1.5、2分钟三组,并与标准雾化装置Dosimeter及雾化器(简称D)进行支气管激发试验同步对比研究。结果显示,三组受试者用两种雾化装置所获测值间地显著性差异(p〉0.2~0.5),且皆呈直线正相关。M与D皆雾化吸入1分钟时,M所需药物深度大于D,相关系数(0

关 键 词:支气管激发试验  压缩机雾化器  哮唏  诊断

Application of model 4650 type I compressor atomizer in bronchial challenge test]
Y Yuan,Y Wang,J Zeng,T He.Application of model 4650 type I compressor atomizer in bronchial challenge test][J].Journal of West China University of Medical Sciences,2000,31(2):262-264.
Authors:Y Yuan  Y Wang  J Zeng  T He
Institution:Department of Internal Medicine, First Affiliated Hospital, WCUMS, Chengdu 610041.
Abstract:This study was directed to the feasibility of applying a simple atomizer-model 4650 type I (abbrev. M)-to bronchial challenge test. 92 cases of asthma were divided into 3 groups randomly. All of them were subjected to the bronchial challenge test by M atomizer, and by standard Dosimeter atomizer (abbrev. D) as a comparison. In the test by M atomizer, the times for inspiring challenging medicine were 1, 1.5 and 2 minutes for the 3 groups respectively, while the time for D atomizer was 1 minute for all. The results showed no significant differences (P > 0.2-0.5) between the two atomizers in the 3 groups, their values were linear correlated. When the inspiring time was 1 minute for both M and D, the test needed a higher concentration of challenging medicine for M than for D, their coefficient of correlation (r = 0.285) was relatively low. When inspiring time postponed to 1.5 minute for M, the difference in medicine concentration between M and D was smallest (-0.075 g/L), r = 0.665. However, a further postponed inspiring time to 2 minute for M reversely broadened their difference. These results indicated that the efficiency of M atomizer was a little lower than that of D, postponing the inspiring time for M could make up this weakness. A 1.5 minute inspiring time for M atomizer was the suggestion. Some modifications on M atomizer were done by us for a better efficiency, and the cheap and popular M atomizer could be a good replacement in bronchial challenge test.
Keywords: