Dosimetry of Sr-90 ophthalmic applicators |
| |
Authors: | C S Reft F T Kuchnir I Rosenberg L C Myrianthopoulos |
| |
Affiliation: | Department of Radiation and Cellular Oncology, University of Chicago, Illinois 60637. |
| |
Abstract: | Sr-90 ophthalmic applicators are commonly used for the treatment of superficial eye disorders. Although a variety of dosimetric devices such as film, thermoluminescent dosimeters (TLD's), ion chambers, and radiochromic foils have been used to measure the peak dose at the applicator surface, there is no internationally agreed upon calibration procedure. Recently, large discrepancies among calibrations of the same applicator at three institutions have been reported. Here we describe a technique to obtain the peak dose rate at the applicator surface using LiF TLD's. The technique can be used for the calibration of flat as well as curved surface applicators. Results for two flat and three concave applicators are presented. Our measurement of the surface dose rate for one of the flat applicators is compared with those obtained by four other institutions, each using different dosimetric devices. |
| |