A New Rate-Modulated Pacemaker System Optimized by Combination of Two Sensors
Authors:ECKHARD ALT  HEINZ THERES  MICHAEL HEINZ  MARKUS MATULA  RUDOLF THILO  HANS BLÖMER
Institution:1st Medical Clinic, Rechts der Isar Medical Center, Technical University of Munich, Federal Republic of Germany.
Abstract:A new rate-modulated pacemaker system optimized by combination of two sensors is described. The parameter body activity and central venous blood temperature control the pacemaker rate. The specific characteristic of each parameter determines its role within the algorithm. While the motion sensor yields a fast reaction following the onset or a change of stress intensity, central venous blood temperature corresponds better to body metabolism. An indication of increased exercise from the motion sensor results in an accordingly rapid increase in the pacing rate. Unless this increased exercise is confirmed by an increase in central venous blood temperature within 2 or 3 minutes, the new motion level will be assumed to be the new baseline motion value and the pace rate will return to a basic pacing rate. Prolonged inappropriate responses are therefore avoided. Longer lasting exercise, fever and nonphysiological signals are recognized and handled safely. Exercise tests with five volunteers under various conditions showed pacing rate behavior that was close to normal.
Keywords:rate-modulated cardiac pacing  temperature pacing  activity pacing