Humidification and heating of anesthetic gases during pediatric anesthesia using the Cicero Anesthesia Workstation] |
| |
Authors: | J M Strauss J Hausd?rfer H Hagemann D Schr?der |
| |
Affiliation: | Abteilung An?sthesie III, Medizinische Hochschule Hannover. |
| |
Abstract: | A series of 52 infants underwent general or urological surgery; all were ventilated with the CICERO. Two different flows of fresh gas were used. In group I (n = 21) the fresh gas flow was set exactly at the level of the minute volume, representing a half-open, non-rebreathing system. In group II (n = 31) the fresh gas flow was adjusted to 10% of the required minute volume. Temperature and relative humidity of the inspired gas were measured continuously close to the tracheal tube. Anaesthesia was accomplished with 2 vol% isoflurane, 21-30 vol% oxygen in nitrous oxide. The results were compared with those achieved with our time-tested paediatric equipment, a SERVO 900D ventilator with a Fisher-Paykel humidifier (Group III, n = 35). Using a high fresh gas flow, no increase in relative humidity in the inspired gas could be detected. The values varied between 12% and 25% (group I). Reducing the flow of fresh gas as indicated above resulted in an increase in the relative humidity (group II). Over the evaluated period of 2 h, humidity increased slowly from an initial mean value of 20% to a maximum of something over 70%. Using the SERVO 900D ventilator combined with the Fisher-Paykel humidifier, humidity reached a value of greater 90% within 10 min after activation of the heated cascade. Humidity in the inspired gas should exceed 70% to avoid damage to infant airways. This will not be attained until after more than 2 h with unaided breathing systems, by when most operations performed on paediatric patients will already be over. Condensed water may aspirated by small infants. This potentially dangerous situation was only encountered in the CICERO circuit, and not in the system protected by the Fisher-Paykel cascade. Dry gases can result in thickened mucous and in obstruction of a small tracheal tube, which requires emergency reintubation. With artificial airways dry gases damage the ciliated epithelia of the trachea and cause loss of water and body heat. The temperature of the "cold" gases varied within a range of 21-33 degrees C and could not be adjusted by the anaesthetist. In the CICERO system, heating the gases at the valve only prevents mechanical failure caused by water condensation. In pediatric anaesthesia, variable heating and non-condensing humidity are essential. The dry and heated gases of the CICERO are not acceptable in the daily practice of paediatric anaesthesia. |
| |
Keywords: | |
|
|